BI-O.K. EO Vials - 100 per box
BI-O.K.™ Ethylene Oxide Biological Indicator vials are easy to use. First, simply place one vial in each sterilization load, either in a pack or test pack. Then, upon completion of the EO cycle, activate the vial by crushing the glass ampoule with the provided vial activator. This releases the spores into the nutrient media, initiating the incubation process.
After incubation at 37°C for 48 hours, the color change of the media indicates whether the cycle was effective. If the media remains red, indicating the absence of spore growth, the sterilization process was successful. However, if the media turns yellow, it signifies spore growth, indicating a failure of the sterilization cycle.
BI-O.K.™ vials adhere to stringent quality standards and are rigorously tested to ensure their reliability in detecting even the most challenging microorganisms.
BI-O.K.™ Ethylene Oxide Biological Indicator Vials Features and Benefits
Self-contained design for convenient operation
Robust polycarbonate vial for durability
Visual spore detection for straightforward interpretation
Easy-to-use vial activator for efficient activation
Distinct test results with visible color change: red indicates a passed cycle, yellow indicates a failed cycle. Be sure to reprocess these instruments before using them.
Specifications:
Storage Conditions: BI-O.K.™ indicator vials are stored under normal room conditions, 59°F-86°F (15°C-30°C) and 40-70% relative humidity. Store away from sterilants and other chemicals.
Shelf Life: 24 months
Gas-Chex® Ethylene Oxide Indicator Strips
Monitor the critical parameters of Ethylene Oxide (EO|EtO) sterilization.